The Ethicon GEN11 Endo-Surgery generator combines advanced technology, multifunctionality and touchscreen simplicity in one sleek and easy-to-operate unit. No matter what the procedure or which Ethicon device you plan to use, the Ethicon Endo-Surgery generator will be the only power source you’ll need. This unit supports Harmonic Ultrasonic and Enseal Advanced Bipolar Devices.

The Ethicon GEN11 is an electrosurgical generator is simple and easy to use. The Ethicon Generator comes with a high-resolution touch screen display that can be seen clearly from a variety of angles. At the same time the generator can recognize when instruments are connected automatically The GEN11 is powered by radio frequencies and is usually used in laparoscopic and gynecological surgeries. The Ethicon generator can support Harmonic Ultrasonic and Enseal Advanced Bipolar devices.

 

  • Universal connector and automatic instrument recognition.
  • Touchscreen for easy setup and operation.
  • High-resolution display with wider viewing angles.
  • Compact design takes up less space in the OR.
  • Software updates available via USB memory stick.

Dimensions (Generator)

  • Height: 7.5 in (13.6 cm)
  • Width: 14.8 in (35.5 cm)
  • Depth: 13.7 in (35 cm)
  • Weight: 13 lbs (5.9 kg)

Dimensions (Cart)

  • Height: 37.5 in (95.3 cm)
  • Width: 22.1 in (56.2 cm)
  • Depth: 18.9 in (48 cm)
  • Weight: 37 lbs (16.8 kg)

Dimensions (Footswitch)

  • Height: 4.1 in (10.4 cm)
  • Width: 13.45 in (34.2 cm)
  • Depth: 7.5 in (19 cm)
  • Weight: 8 lbs (3.6 kg)

Enseal Output

  • Bipolar, no neutral electrode required
  • 100 VAC RMS maximum
  • 135 watts maximum (rated load 15 ohms, unless otherwise specified in the instrument IFU)
  • 300 – 490 kHz (330 kHz, unless otherwise specified in instrument IFU)

Harmonic Output

  • 150 VAC RMS maximum (Unless otherwise specified in the instrument IFU)
  • 35 watts continuous (Unless otherwise specified in the instrument IFU)
  • 30 – 80 kHz (55.5 kHz unless otherwise marked in instrument IFU

Power

  • Main Input: 100–240 V, 50/60 Hz, 500 VA
  • Main Fuses: T5AL 250 V (Time-delay, 5 Amp, glass body, 5x20mm package size, quantity: 2)

Ambient Operating Conditions

  • Temperature: 59°F–80.6°F, 15°C–27°C
  • Humidity: 30%–75% non-condensing
  • Atmospheric Pressure Range: 700 hPa–1060 hPa

Transport and Storage Conditions

  • Temperature: -31°F–129.2°F, -35°C–+54°C
  • Humidity: 10%–95% non-condensing
  • Atmospheric Pressure Range: 700 hPa–1060 hPa

Protections / Classifications

  • Degree of Protection Against Electric Shock: Type CF applied part consisting of the Harmonic or Enseal instruments
  • Class of Protection Against Electric Shock: Class 1
  • Ingress Protection G11 Enclosure: IP21
  • Ingress Protection G11 Fotswitch: IP68